Supporting Data


Medical-Physics-Research-Article

Single Sensor Characterization

 

 

Blue Physics

Test

quantity

Beam Energy

mean

Short-term repeatablity (high dose)

stdev

6 MV

0.22 %

 

 

15 MV

0.25 %

Short-term repeatablity (low dose)

stdev

6 MV

0.50 %

 

 

15 MV

0.58 %

Dose response linearity

RMS

6 MV

0.25 %

Angular dependence

RMS

6 MV

0.78 %

Temperature dependence

 

 

0.012 % per C

Dose rate dependence

RMS

6 MV

0.25 %

 

15 MV

0.38 %

 

 

10 FFF

0.72 %

Measurements done by Yoichi Watanabe and Marcos Feijoo at University of Minneapolis Medical Center on December 19th 2020
parallax background

Output factor for radiosurgery Fields

Measurements done by Jose Luis Dumont at Clinica Radioterapia Reñe on May 20th 2021
parallax background

Output factor for fields smaller than 10cm x 10cm

Measurements done by Jose Luis Dumont at Clinica Radioterapia Reñe on February 12th 2021

PDD´s

Measurements done by Jose Luis Dumont at Clinica de Radioterapia Reñe on June 21st 2021

PDD’s

Measurements done by Luis Brualla and Joan Rosello at ASCIRES on March 27th 2021 and Yoichi Watanabe at University of Minnesota on December 22nd 2021

Profiles of small fields

Measurements done by Luis Brualla and Joan Rosello at ASCIRES on January 17th 2021
parallax background